  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 62
C 150 (E2) 12/2012 page 63/90
With every cleaning method, always use adequate personal safety controls.
Following cleaning, leave the unit door open or remove the access port plugs.
The neutral cleaning agent may cause health problems in contact with skin and if ingested.
Follow the operating instructions and safety hints labeled on the bottle of the neutral clean-
ing agent.
Recommended precautions: To protect the eyes use sealed protective goggles. Suitable protective
gloves with full contact: butyl or nitrile rubber, penetration time >480 minutes.
CAUTION
Contact with skin, ingestion.
Skin and eye damage due to chemical burns.
Do not ingest. Keep away from food and beverages.
Do NOT empty into drains.
Wear protective gloves and goggles.
Avoid skin contact.
Following use of the neutral cleaning agent and prior to hot-air sterilization, remove any agent
residues by using a moistened towel in order to avoid formation of permanent residues.
15.2 Decontamination / chemical disinfection of the CO
2
incubator
Disconnect the CO
2
incubator from the power supply prior to decontamination / disinfection. Pull the pow-
er plug.
You can use the following disinfectants:
Inner chamber
Standard commercial surface disinfectants free from acid or halides.
Alcohol-based solutions.
We recommend using the disinfectant spray Art. No. 1002-0022.
For chemical disinfection, we recommend the disinfectant spray Art. No. 1002-0022.
Any corrosive damage that may arise following use of other disinfectants is excluded from
liability by BINDER GmbH.
With every decontamination / disinfection method, always use adequate personal safety con-
trols.
In case of contamination of the interior by biologically or chemically hazardous material, there are two
possible procedures depending on the type of contamination and charging material.
1. Spray the inner chamber with an appropriate disinfectant.
Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during
the decontamination process.
2. You can sterilize the shelves in a sterilizer or autoclave.