C 150 (E2) 12/2012 page 76/90
18. Troubleshooting
Check connection to power sup-
ply.
Check power supply for voltage of
115V / 208 - 240 V / 230V.
Nominal temperature exceeded
by 10 °C due to unit failure. Over
temperature protective device
(class 1) responds (chap. 9.1).
The miniature fuse for overcur-
rent protection has triggered
(chap. 19.2).
Check unit fuse and replace it if
appropriate. If it responds again,
contact BINDER service.
Message “dor” is displayed.
The outer unit door is not closed
properly.
Close unit door properly.
Chamber does not heat up.
Heating element defective.
Semiconductor relay defective.
Temperature inside too low.
Calibrate and adjust controller.
Temperature inside too low.
Message “tol” is displayed
(temperature tolerance range
alarm)
Wrong controller setting.
Check and, if appropriate, set the
operating function “AL.12” (tem-
perature alarm threshold) (chap.
7.3). Value should be ≥ 1 K.
Temperature inside too high.
Installation site too warm.
Select cooler place of installation
(chap. 3.4).
Difference between the set tem-
perature and the ambient tem-
perature too low.
Difference between the set tem-
perature and the ambient tempera-
ture at least 7°.
Too much external heat load.
Semiconductor relay defective.
Calibrate and adjust controller
Temperature inside too high.
Message “tol” is displayed (tem-
perature tolerance range alarm).
Installation site too warm.
Select cooler place of installation
(chap. 3.4).
Difference between the set tem-
perature and the ambient tem-
perature too low.
Difference between the set tem-
perature and the ambient tempera-
ture at least 7°.
Too much external heat load.
Semiconductor relay defective.
Wrong controller setting.
Check and, if appropriate, set the
operating function “AL.12” (tem-
perature alarm threshold) (chap.
7.3). Value should be ≥ 1 K.