C 150 (E2) 12/2012 page 77/90
Temperature inside too low.
Message “otc” is displayed
(safety controller alarm).
Limit temperature reached. Safe-
ty controller (chap. 9.2) set too
low.
Check the setting of the safety
controller. Choose a suitable limit
value (chap. 9.2).
Temperature inside too high.
Message “otc” is displayed
(safety controller alarm).
Safety controller (chap. 9.2) de-
fective.
Chamber heating permanently,
set point not held.
Semiconductor relay defective.
Message “End” is displayed
Sterilization cycle completed.
Start up the unit (chap. 5).
Pressure alarm. Message “P.Lo”
is displayed (low pressure
alarm).
CO
2
cylinder is not connected
correctly.
Correctly connect the gas cylinder.
Connected gas cylinder is closed
or empty.
Open or replace gas cylinder.
Gas hose is dirty or obstructed.
Check the tube system for dirt
accumulation or obstruction, clean
or replace it.
Pressure sensor system defective
CO
2
concentration inside too
low.
CO
2
cylinder is not connected
correctly.
Correctly connect the gas cylinder.
Connected gas cylinder is closed
or empty.
Open or replace gas cylinder.
Gas hose is dirty or obstructed.
Check the tube system for dirt
accumulation or obstruction, clean
or replace it.
CO
2
concentration inside too
low. Message “con” is displayed
(CO
2
tolerance range alarm)
Wrong controller setting.
Check and, if appropriate, set the
operating function “AL.2” (CO
2
alarm threshold) (chap. 7.3). Value
should be ≥ 1.0 vol.-%.CO
2
.
CO
2
concentration inside too
high.
CO
2
concentration inside too
high. Message “con” is dis-
played (CO
2
tolerance range
alarm)
Wrong controller setting.
Check and, if appropriate, set the
operating function “AL.2” (CO
2
alarm threshold) (chap. 7.3). Value
should be ≥ 1.0 vol.-%.CO
2
.
Actual value of CO
2
deviates
largely compared with a refer-
ence method.
The pH indicator of the cell me-
dium changes its normal color
CO
2
sensor system defective.
Transfer the cultures to another
incubator and contact BINDER
Service.
Recovery time (up to 5 vol.-%
CO
2
) after doors were open for 2
minutes is < 2 minutes.
CO
2
sensor system defective.
Recovery time (up to 5 vol.-%
CO
2
) after doors were open for 2
minutes is > 10 minutes.
Check gas supply (cylinder, con-
nections, hose system).
Insufficient CO
2
input pressure.
Gas fine filter obstructed.