C 150 (E2) 12/2012 page 81/90
C 150-UL electrical connection data (for the USA and Canada)
Electrical Data
IP system of protection acc. to EN 60529
IP
20
Nominal voltage (±10 %) 50/60 Hz
V
115 1N~
Nominal power
kW
1.40
Energy consumption 2) at 37 °C/ 98.6°F
Wh/h
120
Power plug
NEMA
5-20P
Installation category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Unit fuse
6.3 X 32 mm / 250V / super-time-lag TT / 16A
Legend:
1) to 98% of the set value
The recovery times of the gas concentrations inside the chamber following the door being opened co-
incide with a connection pressure of 2.0 bar / 29 psi. Decreasing supply pressure results in longer re-
covery times.
2) Use this value for sizing air condition systems.
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77°F and a power supply voltage fluctuation of ±10. The temperature data is determined in ac-
cordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances
of 10% of the height, width and depth of the inner chamber.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
19.5 Important conversion data for non-SI units
1 ft = 0.305 m = 0.000305 km
1 m = 100 cm = 3.28 ft = 39.37 inch
1 km = 1000 m = 3280.83 ft
1 mbar = 0.0145 psi
19.6 Conversion table for gas inlet pressures, bar psi
bar
psi
bar
psi
bar
psi
1
14.5
3
43.5
5
72.5
1.5
21.7
3.5
50.7
5.5
79.7
2
29.0
4
58.0
6
87.0
2.5
36.3
4.5
65.2