4.5 Electrical connection
The CO
2
incubator C 150 has a fixed power supply connection cable 1800 mm / 70.87 inches in length
and a shock-proof plug. Voltage 230 V (1N~) +/- 10%, 50/60 Hz
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (centrally located at the bottom of the left-hand side of the unit,
see chap. 1.4)
When connecting, please observe the regulations specified by the local electricity supply company and
as well as the VDE directives (for Germany)
Pollution degree (acc. to IEC 61010-1): 2
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 19.4).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in
case of danger.