C 150 (E2) 12/2012 page 64/90
In case of eye contact, the disinfectant spray may cause eye damage due to chemical
burns. Follow the operating instructions and safety hints labeled on the bottle of the disin-
fectant spray.
Recommended precautions: To protect the eyes use sealed protective goggles.
VORSICHT
Eye contact.
Eye damage due to chemical burns.
Do NOT empty into drains.
Wear protective goggles.
Following frequent use of the disinfectant spray and prior to hot-air sterilization, remove any
agent remainder by using the neutral cleaning agent and then a moistened towel to avoid for-
mation of permanent residues.
After using the disinfectant spray, allow the incubator to dry thoroughly, and aerate it suffi-
ciently.
15.3 Disinfection of the CO
2
sensor
To ensure complete disinfection and proper function of the sensor, BINDER recommends a wipe disinfec-
tion of the sensor head with pure alcohol or non-corrosive alcohol-based surface disinfectants. The disin-
fectant must be non-corrosive and free of chlorine or any acid. We recommend using the disinfectant Art.
No. 1002-0022. Avoid strong shocks when handling the CO
2
sensor.
CAUTION
Excess temperature.
Immersion of sensor into liquids.
Shocks of the CO
2
sensor.
Damage to the CO
2
sensor.
Do NOT immerse the CO
2
sensor into liquids.
Do NOT expose the CO
2
sensor to autoclaving.
Do NOT expose the CO
2
sensor to hot-air sterilization.
Avoid strong shocks of the CO
2
sensor (by putting it down hard, or dropping).
We recommend regular disinfection of the CO
2
sensor.
CAUTION
Connecting or removing the CO
2
sensor during operation.
Damage to the CO
2
sensor.
Connect or remove the CO
2
sensor only with the unit turned off.