4.2 Electrical connection
The refrigerated incubator comes with an IEC connector plug.
Model
Art. no.
(x = 0 or 1)
Power plug
Voltage
+/-10 %
Current
type
Power
frequency
Unit fuse
KT 53
KT 115
9x20-0250
9x20-0209
Shock-proof
plug
200 V to 240 V 1N~ 50/60 Hz 10 A
KT 53-UL
KT 115-UL
9x20-0255
9x20-0227
NEMA 5-15P 100 V to 120 V 1N~ 50/60 Hz 10 A
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (left unit side, bottom right-hand, see chap. 1.4).
When connecting, please observe the regulations specified by the local electricity supply company as
well as the VDE directives (for Germany).
Pollution degree (acc. to IEC 61010-1): 2
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 23.4).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in
case of danger.