Binder KT 53 User Manual Page 27

  • Download
  • Add to my manuals
  • Print
  • Page
    / 135
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 26
KT (E6) 09/2013 Page 27/135
6.2 Operating modes
In the “control off” mode (chap. 6.2.1), the controller is non-functional and displays only the actual
values. There is no heating or refrigeration. The temperature approximates the ambient value, the fan
turns with 40 % speed.
You can enter the desired set point values in fixed value” mode (chap. 8). The controller then operates
as a fixed-point control, i.e., it reaches and maintains the defined temperature set-point until the next
manual change.
The T4.12 program controller also permits running a time program (chap. 9) or a week program (chap.
10). You can program temperature cycles and define also the fan speed for each program section. The
controller offers 52 time program places with up to 100 sections each. The week program mode offers 8
week program places with up to 30 shift points for each week program.
6.2.1 Activating the “control off mode or change to “fixed value” operating mode
To select the “control offor “fixed value” operating mode, go to Menu > Controller mode
Fixed value
08.03.2013 05:05:06
General menu.
Select “Controller Mode”
and press the operating button.
..\ Menu
Controller mode
Event list
Alarms
Setpoints
Safety controller
Programs
Import/Export
Settings
Close Home
Fixed value
08.03.2013 05:05:06
Submenu “Controller Mode”.
Select the desired controller mode
“Control off” or “Fixed value”
and press the operating button.
..\ Controller mode
Control off
Fixed value
Close Home
Page view 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 134 135

Comments to this Manuals

No comments