19.8 Water protected internal socket (option)
You can turn on and off the voltage of the interior socket via the controller menu (chap. 7.2).
The internal socket is splash proof.
IP system of protection 67, 100-240 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 50 °C / 122°F.
WARNING
Exceeding the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
Do NOT exceed the temperature set-point of 50 °C / 122°F.
Set the overtemperature safety controller to “Limit” and a value of 50 °C / 122°F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
range.
CAUTION
Risk of short circuit.
Damage to the unit.
Use the supplied plug only (IP protection type 67
). Plug it in and turn it to secure
contact.
If the socket is not used, close the lift-up lid and turn it to secure.