11. Troubleshooting
Set-point temperature is not
reached after specified time.
Unit door not properly closed.
Completely close unit door.
Controller not adjusted, or
adjustment interval exceeded.
Calibrate and adjust controller.
Wrong voltage.
Check power supply for voltage of
115V or 230V.
The fan doesn’t turn or turns too
slowly.
Set the fan speed to 100%.
Chamber heating permanently,
set-point not held.
Contact BINDER service.
Semiconductor relay defective
Controller not adjusted, or
adjustment interval exceeded.
Calibrate and adjust controller.
Red heating control light in the
Heating element defective.
Contact BINDER service.
Semiconductor relay defective.
Chamber doesn’t heat up.
Red heating control light in the
display is not lit.
Controller display working.
Timer run off.
Program the timer or change to time
function Continuous operation
Semiconductor relay defective.
Contact BINDER service.
Unit without function, only the
green “stand-by” LED is lit
Unit in stand-by mode.
Press down the ON/OFF button (5)
until the display lights up.
Temperature inside the chamber
too high, Red alarm pilot lamp of
safety device (7a) is lit.
Safety device class 3.1 has
responded.
Check the settings of the
temperature set-point and of the
safety device class 3.1 (chap. 7).
Unit without any function.
Check connection to power supply.
Unit fuse has responded.
Check unit fuse and replace it if
appropriate. If it responds again,
Controller defective. Contact BINDER service.
Deviations from the indicated
heating-up times.
Oven fully loaded.
Charge the oven less or consider
longer heating-up times.
Message „1999” in the controller
Sensor rupture between
Contact BINDER service.
The controller returns to Normal
Display from any level.
No button was hit for more
than approx. 30 sec.
Repeat entries, enter the values
rapidly.
Only qualified service personnel authorized by BINDER must perform repair.