8.6 Water protected internal socket (option)
You can turn on or off the disconnectable water protected internal socket by switch (9), independent of the
incubator operating or not. Thus, devices operated inside the incubator can be started or stopped without
any need to open the chamber doors.
The internal socket is splash proof.
IP system of protection 67 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 90 °C / 194 °F.
WARNING
Exceeding of the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
Do NOT exceed the temperature set-point of 90 °C / 194 °F.
Set the mechanical thermostat class 3.1 to 90 °C / 194 °F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
range.
CAUTION
Risk of short circuit.
Damage to the unit.
Use the delivered plug only (IP protection type 67). Plug it in and tighten it by screwing it
to secure contact.
If the socket is not used, close the screw lid and turn it to secure.
Controller shutdown by the On/Off switch (5) has no effect on the internal socket.
WARNING
Internal socket switched on even though controller is shut down.
Electrical hazard.
Switch off the internal socket separately, when not in use, by switch (9).