1. Press down
button for approx. 5 seconds.
The display alternately shows unitand the temperature unit:
2. Press again button
.
The display alternately shows rASdand the actual setting of the set-point gradient:
3. Set the desired ramp gradient with buttons
(set-point gradient in °F or °C acc. to setting
in chap. 6.4.1).
4. The set value is automatically adopted after 2 seconds.
During ramp operation the actual set-point (SPr) continually rises in accordance to the entered gradient
from the previously entered set-point to the new one (SP). The actual value follows the set-point value.
About set-point display during ramp operation see chap. 6.2.
6.4.3 Chamber addressing
If several incubators with forced convection BF are networked with a PC via the APT-COM
communication software (option, chap. 8.2), each unit must be allocated a unique address. Addressing
takes place on the chamber controller as follows:
1. Press down
button for approx. 5 seconds.
The display alternately shows unitand the temperature unit:
2. Press again button
.
The display alternately shows rASd and the set-point gradient:
3. Press again button
.
The display alternately shows Adrand the actual setting of the unit address: