Binder KB 23 User Manual Page 44

  • Download
  • Add to my manuals
  • Print
  • Page
    / 101
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 43
KB (E3.1 + E5.1) 11/2014 page 44/100
10. User level
In this menu the following parameters can be se (in brackets the corresponding abbreviated information
given in display 2):
Unit address (Address)
Setting of controller address (1 to 254) for operation with the communication software APT-COM™.
User code (User-cod)
Modification of the user code setting (factory setting 0001”) for access to the user level and the
program editor.
Keep in mind any modification of the user code. Without a valid user code it is no longer
possible to access these levels.
Safety controller mode (Saf.mode)
The set-point mode can be entered as ...
Limit: Maximally permitted temperature as absolute value (example: temperature set-point 37 °C /
98.6°F, limit setting to 39 °C / 102.2°F)
Offset: Maximally permitted temperature exceeding of the actual set-point temperature (e.g. 2 °C).
This maximally permitted temperature automatically changes with every change of the temperature
set-point.
Safety controller set-point (Saf.setp)
The temperature set-point setting of the safety controller (over temperature safety device class 3.1) is
shown and can be changed. An absolute value (e.g. 40 °C / 104°F) in case of setting limit” or a
relative value (e.g. 2 °C) in case of setting Offsetcan be entered.
Regularly check the setting of the set-point mode and of the over temperature safety
device set-point entered in this mode for the temperature value in Fixed value operation or
for the maximum temperature value of the selected temperature program in Program
operation mode.
Decimal point position (Decimal)
Selection if integer values or one position after the decimal point can be entered. The integer
representation is shown in Display 2 (set-point entry) while the actual value in Display 1 is always
displayed with one decimal point.
Audio Alert (Buzzer)
Inactive: no audible signal (buzzer) in case of an alarm event.
Active: in case of an alarm event (see chap. 13.2) an audible signal (buzzer) will sound. It can be reset
by pressing the “EXIT button.
Selection of controller menu language (Language)
German, English, or French can be selected.
Counter of operating hours (Oper.hs)
Information about the number of operating hours currently reached or since the last reset. (no setting,
display only).
Max. number of operating hours (Op.limit)
Entry of a limit number of operating hours, i.e., the maximum number of operating hours that can be
run. Maximum setting: 9999. Reaching the limit has no effect.
Reset operating hours (Op.back)
Reset operating hours to zero.