19. Technical description
19.1 Factory calibration and adjustment
This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using
standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certi-
fied since December 1996 by TÜV CERT). All test equipment used is subject to the administration of
measurement and test equipment that is also a constituent part of the BINDER QM DIN EN ISO 9001
systems. They are controlled and calibrated to a DKD-Standard at regular intervals.
Factory adjustment was done in the center of the usable volume and under vacuum conditions. The sen-
sor is fixed in the middle of the expansion rack in a way ensuring good thermal conductivity (heat conduc-
tion). Measuring is performed in equilibrated state.
CAUTION
Invalid calibration.
∅ Do NOT change between aluminum and stainless steel racks.
Use the delivered expansion racks only.
19.2 VDL technical data
Vacuum oven
Vacuum module (option)
Vacuum oven with op-
tion vacuum module
Plus door handle, connections (depth)
Wall clearance at the back
Wall clearance at the side
Number of expansion racks (aluminum)
Distance between the racks
Usable space per rack (width x depth)
mm / inch
234 x 280 /
349 x 320 /
455 x 440 /
Permissible load per rack