MK (E2.1) 10/2014 page 40/73
9.5 Repetition of a section or several sections within a program
Here we use the example of a set-point ramp temperature program of chap. 9.3. The shaded sections 02
and 03 shall be repeated e.g. 30 times.
Program
Set-point
Fan Section
Operation
Target
No. of
Min.
Max.
The following table shows the program that results, whereby the differences to the table above are shaded.
Sections 02 and 03 will be executed in total 31 times; only then will the program continue.
Entry of the values into the display program table:
To have sections repeated infinitely, enter the number of cycles “Cy” as “-1”.
9.6 Performance after power failure in Program Mode
The program is resumed at the point where the interruption occurred with the latest set-points reached
during the program run. The power failure is noted in the event list. No error message is displayed indicat-
ing that a power failure had taken place.