Any solvent contained in the charging material must not be explosive or inflammable. I.e., irrespective of
the solvent concentration in the steam room, NO explosive mixture with air must form. The temperature
inside the chamber must lie below the flash point or below the sublimation point of the charging material.
Familiarize yourself with the physical and chemical properties of the charging material, as well as the con-
tained moisture constituent and its behavior with the addition of heat energy.
Familiarize yourself with any potential health risks caused by the charging material, the contained moisture
constituent or by reaction products that may arise during the temperature process. Take adequate
measures to exclude such risks prior to putting the alternating climate chamber into operation.
DANGER
Electrical hazard.
Danger of death.
The unit must NOT become wet during operation or maintenance.
The alternating climate chambers were produced in accordance with VDE regulations and were routinely
tested in accordance to VDE 0411 (IEC 61010-1).
CAUTION
The inner chamber, the door window and the access ports will become hot during
operation.
Danger of burning.
Do NOT touch the inner surfaces, the door window, the access ports, or the charging
material during operation.
WARNING
Stability hazard.
Danger of injury.
Damage to the unit and the charging material.
Housing cover breakaway.
Do NOT climb the lower housing cover.
Do NOT load the lower housing cover with heavy objects while the unit door is open.